Electrical Specifications | |||
Precision and phase shift | Range | 100mV/A | 10mV/A |
Current range | 50mA~10A peak | 1A~100A peak | |
Output signal | 3%±5mV | 500m~40A peak:4% 100A(max.)15% | |
Frequency range(-3dB) | DC~100kHz | ||
Phase shift | DC~65Hz:<1.5 | DC~65Hz:<1° | |
Load impedance | ≥1MΩ and ≤100pF | ||
Insert impedance | 0.01Ω | ||
Noise | 6mV | 600μV | |
Slew Rate | 0.3V/μS | 20mV/μs | |
Rise/fall time | 3μs | 4μs | |
Measurement conditions:23°C±5°C,20~75%RH,Sinusoidal frequency:48~65Hz,External magnetic field <40A/m,no load flow, the center of the test sample,load impedance:1MΩ | |||
Working voltage | Max 600Vrms | ||
Common mode voltage | Max 600Vrms | ||
Battery | 9V Alkaline battery (meet up with: NEDA 1604A,IEC 6LR61) | ||
Low battery indicator | >6.5V,LED is green | ||
Overload indicator | Red LED means the measure current is too big and it need to change the measuring range | ||
Mechanical index | |||
Oper.Temp | 0~+50°C | ||
Preservation temperature | -3~+80°C | ||
Temperature effect | Per 10°C,<0.2% | ||
Operation relative humidity | +10°C~+30°C:85±5% elative humidity(non-condensing): +40°C~+50°C:45±5% elative humidity(non-condensing) | ||
Operating height | 0~2000m | ||
Maximum EZhong insert | Dia.11.8mm | ||
Dimension | 231х36х67mm | ||
Weight | 330g(have cell) | ||
Output | 2 meters coaxial cable, terminal with insulation BNC connector jack |
*From: |
Enter your email |
*Message: |
Enter your Message |