Summary of Source and Measure Functions | ||||
Function | Measure | Source | ||
DC V | 0~50V(the upper display ± 30V) | 0~10V | ||
DC mA | 0~24mA(the upper display ± 24mA) | 0~24mA | ||
Frequency | 1.000Hz~100.00kHz | 0.00Hz~20.000kHz | ||
Resistance | 0~3200Ω | 0~3200Ω | ||
DC mV | 0~100mV(the upper display±200mV) | 0~100mV | ||
RTD | Pt100,Pt1000,Cu50,Cu100 | Pt100,Pt1000,Cu50,Cu100 | ||
TC | E,J,K,T,B,R,S,N | E,J,K,T,B,R,S,N | ||
Others | Loop supply,Step,Ramp,Dual display | |||
DC Voltage Measurement | ||||
Range | Maximum measurement range | Resolution | Accuracy(%of reading+counts) | |
28V(upper display) | -33V~33V | 0.001V | 0.025+2 | |
200mV(upper display) | -200mV~200mV | 0.01mV | 0.025+2 | |
50V(lower display) | -1V~60V | 0.001V | 0.02+2 | |
100mV(lower display) | -15mV~125mV | 0.01mV | 0.02+2 | |
DC mA measurement | ||||
Range | Maximum measurement range | Resolution | Accuracy(%of reading+counts) | |
20mA(upper display) | -24mA~24mA | 0.001mA | 0.025+2 | |
20mA(loop of upper display) | 0~24mA | 0.001mA | 0.025+2 | |
20mA(lower display) | 0~24mA | 0.001mA | 0.02+2 | |
Resistance Measurement (upper display) | ||||
Range | Maximum output range | Resolution | Accuracy(%of reading+counts) | |
400Ω | 0~440Ω | 0.1Ω | 0.05+2 | |
3200Ω | 420Ω~3300Ω | 1Ω | ||
On-off test | 0~200Ω | 1Ω | 0.05+1 | |
Resistance measurement (lower display) | ||||
Range | Maximum measurement range | Resolution | Accuracy(Ω) | |
Two-wire, three-wire | Four-wire | |||
400Ω | 0~440Ω | 0.01Ω | 0.15 | 0.1 |
3200Ω | 420Ω~3600Ω | 0.1Ω | 1 | 0.5 |
*From: |